Products:
Workstation
|
Workstation
Applications:
Hospital and medical laboratories
Pathology procedures
Industrial processes and
production lines
Drug and chemical compounding
Asbestos testing
The HS-4500 is a completely self contained, filtered-air workstation and work surface in one package. The angled clear acrylic shield over the work
area provides full visibility with easy access to instruments or containers
inside. The built-in gooseneck worklight is adjustable to provide illumination
at the most effective angles. This unit is often selected for use with a
microscope and is available with an appropriate opening in the face shield
to accommodate most stereo zoom microscopes. The filter housing is designed
to hold one or more filters simultaneously, allowing one station to be used
for multiple pollutants or procedures. Filters are easily accessed for replacement
with no tools required. The system is powered by a 530 cfm rated blower
and it has a variable speed control for quiet and efficient operation. There
is a built in AC outlet for plugging in a microscope, soldering iron, dictation
equipment, additional lighting or other electrical device. The workstation
weighs approximately 65 pounds and it can be conveniently moved to any workplace.
Workstation Options:
Airflow Alarm Unit
Blocked Filter Alarm Unit
Acrylic Polish - Anti Static Surface Cleaner
Specifications |
Model No: HS-4500S
- solid acrylic panel
Model No: HS-4500C
- cutout for microscope
Overall Size: 40.5"W x 16"D x 19"H
Workspace Dim.: 25"W x 16"D x 15.75"H
Power requirements: 115V AC, 100 Watts, 50/60 Hz, 230V
AC optional for export
Weight: 65 lbs.
Blower: 530 CFM, (free air), Backward Curved Impeller
Noise Level: A weighted noise level 68dB at 1 meter |
|
Airflow Performance Statistics
|
Filters Installed |
Max Airflow
CFM
|
Face Velocity
FPM
|
None |
295
|
310
|
HEPA |
204
|
215
|
HEPA/Dacron/.5" Carbon |
180
|
190
|
HEPA/Dacron/2" Carbon |
90
|
95
|
0.5" Carbon |
61
|
275
|
2" Carbon |
119
|
125
|
2 each - 2" Carbon |
83
|
87
|
All numbers refer to average readings taken at full blower speed
|
Please Note: We can manufacture this unit in many different sizes and/or configurations. Please contact us to discuss your application or send us your specifications. Custom Design Specifications Worksheet
NOTE: The above units are supplied without filters. Selection of filters is dependent upon application. Click here for information on Filters
|
 |